If you recognize persistent film in the measuring cell of your EasyDens clean your EasyDens with a suitable laboratory cleaner like:
- Mucasol® (3 % solution)
- Desana® (diluted 300 mL in 10 L water)
- Alconox® (as 1 % solution)
- Tergazyme® (as 1 % solution)
- Ethanol (96 % v/v)
Be sure to attend to the product information concerning the cleaning agent’s concentration suggested by the manufacturer.
If the deposits are very stubborn, let the cleaning solution soak in the cell for about 20 minutes.
After cleaning with a laboratory cleaner always rinse thoroughly with distilled water.
Note: Do not use aggressive solvents (e.g. Acetone) to clean the device. Only use approved and compatible cleaning agents, otherwise your device could suffer permanent damage.